Medical Policy
Subject: Leadless Pacemaker
Document #: SURG.00150Publish Date: 09/27/2023
Status: ReviewedLast Review Date: 02/16/2023
Description/Scope

This document addresses a single chamber implantable transcatheter pacing system to monitor and regulate the heart rate and rate-responsive bradycardia.

Position Statement

Investigational and Not Medically Necessary:

Use of the leadless pacemaker is considered investigational and not medically necessary for all applications.

Rationale

The Medtronic Micra Model MC1VR01 single chamber implantable transcatheter pacing system (TPS) with SureScan technology (Medtronic, Inc., Mounds View, MN) is an MR (Magnetic Resonance) conditional programmable cardiac device that monitors and regulates the heart rate by providing rate-responsive bradycardia pacing to the right ventricle. The device senses the electrical activity of the heart, using the sensing and pacing electrodes on the titanium capsule of the device. It monitors the heart rhythm for bradycardia and responds by providing pacing therapy, based on the pacing parameters programmed into the device. The device provides rate response, controlled through an activity-based sensor, and also provides diagnostic and monitoring information for guidance in the pacing system evaluation. 

The Micra TPS contains a one inch-long, self-contained pacemaker that is implanted directly into the right ventricle chamber of the heart. It works like other pacemakers to control the heartbeat but, unlike other pacemakers, the Micra TPS does not have leads and does not require a subcutaneous pocket. The Micra TPS is a single-chamber pacing system, which paces only the right ventricle of the heart. The Micra TPS is implanted percutaneously directly into the heart via the femoral vein using a catheter delivery system and is attached to the right ventricle with four small nictinol prongs (tines). According to the Medtronic Clinician manual for the Micra TPS device:

Patients with an implanted Micra Model MC1VR01 pacing system can safely undergo an MRI scan provided the system meets the requirements described in the Medtronic MRI Technical Manual.

The FDA clearance of this device was based on an evaluation of data from a single study. This was a prospective, multicenter, single-arm study of 719 subjects implanted with the Micra TPS with two primary outcomes: efficacy and safety. Enrolled subjects met either class I or II indications for pacing and were candidates for single-chamber pacing. The majority (64.0%) had bradycardia associated with persistent or permanent atrial tachyarrhythmia, 17.5% had sinus-node dysfunction, 14.8% had atrioventricular (AV) block. A planned interim analysis was completed when 300 subjects reached 6 months of follow-up. The primary efficacy endpoint, the percent of subjects with low and stable pacing capture thresholds at 6 months, was 98.3% (95% confidence interval [CI], 96.1-99.5; p<0.001). The primary safety endpoint, freedom from system-related or procedure-related major complications, was 96.0% (95% CI, 93.9-97.3; p<0.001). Additionally, safety outcomes were compared to historic controls from six previous transvenous pacemaker trials. While there were significant differences between the study and control subjects, the implanted study group experienced fewer hospitalizations (2.3% vs. 3.9%) and fewer system revisions (0.4% vs. 3.5%). This clinical trial will continue to follow subjects for at least an additional 12 months to evaluate the long-term performance of the Micra TPS. The evidence from this study is considered preliminary and insufficient to demonstrate the long-term safety and efficacy of the Micra TPS, as compared to conventional pacemaker devices. Further study is needed with longer term outcomes data (Reynolds, 2016).

Small studies of the Micra TPS and a similar device, the Nanostim Leadless Cardiac Pacemaker (LCP) (St. Jude Medical, St. Paul, MN, acquired by Abbott Laboratories, Abbott Park, IL) have reported some favorable short-term results for individuals requiring single chamber pacing. A medical device advisory by St. Jude Medical for the Nanostim was issued on October 11, 2016, due to premature battery depletion issues encountered with the device. Implantation of the Nanostim device has been halted at the present time, and the Nanostim has not been cleared by the FDA, to date. Registry data on the Micra TPS is being collected by the manufacturer in the Micra TPS Post-approval registry, which is intended to enroll 2450 subjects implanted with the Micra TPS in a global, prospective, observational, multi-site registry (NCT02536118). Registry participants will be prospectively followed for a minimum of 9 years post-implant or until registry closure. Enrolled individuals will have scheduled follow-up visits at least annually, or as prompted by reportable adverse events, with a total estimated registry duration of 11 years. The current status of the registry is active, not recruiting participants and the estimated completion date is August, 2026.

Background/Overview

The Micra Transcatheter Pacing System (TPS) (Pacemaker Model MC1VR01 and Programmer Application Software Model SW022 Version 1.1) was granted expedited priority review and clearance by the U.S. Food and Drug Administration (FDA) on April 6, 2016 for use in individuals who have slow or irregular heart rhythms and who may benefit from a single-chamber pacemaker system, or in whom placement of a traditional system is difficult. Approval was based primarily on the results of a single-arm pivotal trial, which will continue to follow trial subjects for at least 12 months to further evaluate the long-term performance of the device.

The approved FDA indications for the Micra TPS are as follows:

The Micra Transcatheter Pacing System is indicated for use in patients who have experienced one or more of the following conditions:

According to the FDA labeling, the Micra Model MC1VR01 pacemaker is contraindicated for individuals who already have the following types of devices implanted:

In addition, the device is contraindicated for persons with any of the following conditions:

Also, the Micra TPS is not for use in individuals who can’t tolerate a single dose of 1.0 mg dexamethasone acetate. Magnetic resonance imaging (MRI) may be contraindicated for some individuals with an implanted Micra TPS device. The Micra TPS is currently the only leadless pacemaker that is FDA cleared for marketing in the United States. In January 2020, the Micra AV TPS (model MC1AVR1) and application software model (SW044) were cleared as premarket approval (PMA) supplements to the original Micra TPS device. The Micra AV includes an enhanced algorithm for AV synchronous pacing.

On March 31, 2022 the FDA cleared the Aveir Leadless Pacemaker System (Abbot Medical, Sylmar, CA) which contains the following components:

The Aveir Leadless Pacemaker (LP) is indicated for individuals with bradycardia and:

Rate-Modulated Pacing is indicated for individuals with chronotropic incompetence, and for those who would benefit from increased stimulation rates concurrent with physical activity.

The MR Conditional Aveir LP is conditionally safe for use in the MRI environment according to the instructions in the Abbott MRI-Ready Leadless System Manual.

The Aveir Delivery Catheter is intended to be used in the peripheral vasculature and the cardiovascular system to deliver and manipulate an LP. Delivery and manipulation include implanting an LP within the target chamber of the heart.

The Aveir Link Module is intended to be used in conjunction with a Merlin PCS Programmer to interrogate and program an Aveir LP and to monitor LP function during an implant, retrieval, or follow-up procedure.

Sensing and pacing occur between a distal electrode near the helix and the external can of the LP. The LP's proximal end has a feature for docking to delivery and retrieval catheters, providing for repositioning and retrieval capability. The LP communicates bi-directionally with the programmer system via electrical signals conducted between the implanted LP's electrodes and skin electrodes applied to the chest and connected to the programmer system. Consequently, the LP transmits signals using circuits and electrodes already provided for pacing, with data encoded in pulses delivered during the refractory period of the ventricle.
The FDA PMA decision was based on data from the global LEADLESS II phase 2 investigational device exemption (IDE) study that evaluated the Aveir VR in subjects with certain abnormal heart rhythms. The results showed the device met its pre-specified primary endpoints. A subject was considered to have met the confirmatory effectiveness endpoint if the pacing threshold voltage is ≤ 2.0 V at 0.4 ms and the sensed R-wave amplitude is either ≥ 5.0 mV at the 6-week visit or ≥ the value at implant (Han, 2022; Reddy, 2022).

According to the FDA (SSED, 2022):

Phase 1 of this IDE study evaluated the safety and effectiveness of the Nanostim Leadless Pacemaker in a population indicated for a VVI(R) pacemaker. Since the Nanostim Leadless Pacemaker was modified prior to market release and renamed the Aveir Leadless pacemaker, Phase 2 of this IDE study confirmed the safety and effectiveness of these modifications in the Aveir Leadless Pacemaker.

Notably, according to the FDA, “The Aveir Leadless System has not been marketed in the United States or any foreign country.”

On January 18, 2017, the Centers for Medicare and Medicaid Services (CMS) issued a final Decision Memo regarding its national coverage analysis (NCA) for Leadless Pacemakers (CAG-00448N). The memo indicates that CMS will only cover the leadless pacemaker, in accordance with the FDA approved label, when furnished as part of CMS-approved post-approval studies or prospective longitudinal studies meeting criteria, through its Coverage with Evidence Development (CED) program. This CMS Decision Memo was issued after considering public comments submitted during two separate 30-day periods. All clinical research study protocols must be reviewed and approved by CMS. The memo includes a preliminary draft of Medicare's National Coverage Determination (NCD) for leadless pacemakers; the effective date is not yet set. According to the CMS NCA, the following summary is provided:

These devices have not been tested in broader, long-term studies that include real world practice settings and are currently being followed in FDA-required post-market studies…CMS believes that the evidence is promising and sufficient for coverage of leadless pacemakers when furnished in CMS-approved studies under CED. CMS believes that the available evidence is not sufficient to determine long-term health outcomes or to identify the characteristics of the patient, practitioner or facility that predict which beneficiaries are more likely to experience overall benefit or harm from leadless pacemakers. Significant questions remain regarding the potential for deterioration in left ventricular function and other long-term outcomes, as well as device longevity (CMS, 2017).

Definitions

Arrhythmia (or dysrhythmia): Problems that affect the electrical system of the heart muscle, producing abnormal heart rhythms and may be classified as either atrial or ventricular, depending on which part of the heart they originate from.

Atrial Fibrillation (AF): A condition in which the atrium (the heart’s two upper chambers) produce uncoordinated electrical signals.

Bradycardia: A condition where the individual has a slow heart rate, typically defined as a heart rate of under 60 beats per minute (BPM) in adults.

Congestive Heart Failure (CHF), also referred to as Heart Failure (HF): A condition in which the heart can't pump enough blood to the body's other organs. The "failing" heart keeps working but not as efficiently as it should. As blood flow out of the heart slows, blood returning to the heart through the veins backs up, causing congestion in the tissues.

Heart Block: A disease or inherited condition that causes a fault within the heart's natural pacemaker, due to some kind of obstruction (or "block") in the electrical conduction system of the heart. Blockages are classified based on where the blockage occurs - the SA node ("Sinoatrial block"), AV node ("AV block" or AVB), and at or below the bundle of His ("Intra-Hisian" or "Infra-Hisian block" respectively). In severe cases where the heart's ability to control and trigger heartbeats may be completely ineffective or unreliable, heart block can usually be treated by inserting an artificial pacemaker that provides correct electrical impulses to trigger heart beats, compensating for the natural pacemaker's unreliability.

Sinus Node Dysfunction (also called Sick sinus syndrome or sinoatrial node disease): A group of abnormal heart rhythms (arrhythmias) presumably caused by a malfunction of the sinus node, the heart's primary pacemaker.

Sinus Tachycardia: A sinus rhythm (emanating from the SA node) with an elevated rate of impulses, defined as a rate greater than 100 BPM in an average adult.

Coding

The following codes for treatments and procedures applicable to this document are included below for informational purposes. Inclusion or exclusion of a procedure, diagnosis or device code(s) does not constitute or imply member coverage or provider reimbursement policy. Please refer to the member's contract benefits in effect at the time of service to determine coverage or non-coverage of these services as it applies to an individual member.

When services are Investigational and Not Medically Necessary:
For the following procedure codes; or when the code describes a procedure indicated in the Position Statement section as investigational and not medically necessary.

CPT

 

33274

Transcatheter insertion or replacement of permanent leadless pacemaker, right ventricular, including imaging guidance (eg, fluoroscopy, venous ultrasound, ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed

33275

Transcatheter removal of permanent leadless pacemaker, right ventricular, including imaging guidance (eg, fluoroscopy, venous ultrasound, ventriculography, femoral venography), when performed

0795T

Transcatheter insertion of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed; complete system (ie, right atrial and right ventricular pacemaker components)

0796T

Transcatheter insertion of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed; right atrial pacemaker component (when an existing right ventricular single leadless pacemaker exists to create a dual-chamber leadless
pacemaker system)

0797T

Transcatheter insertion of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed; right ventricular pacemaker component (when part of a dual-chamber leadless pacemaker system)

0798T

Transcatheter removal of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography), when performed; complete system (ie, right atrial and right ventricular pacemaker components)

0799T

Transcatheter removal of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography), when performed; right atrial pacemaker component

0800T

Transcatheter removal of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography), when performed; right ventricular pacemaker component (when part of a dual-chamber leadless pacemaker system)

0801T

Transcatheter removal and replacement of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed; dual-chamber system (ie, right atrial and right ventricular pacemaker components)

0802T

Transcatheter removal and replacement of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed; right atrial pacemaker component

0803T

Transcatheter removal and replacement of permanent dual-chamber leadless pacemaker, including imaging guidance (eg, fluoroscopy, venous ultrasound, right atrial angiography, right ventriculography, femoral venography) and device evaluation (eg, interrogation or programming), when performed; right ventricular pacemaker component (when part of a dual-chamber leadless pacemaker system)

0804T

Programming device evaluation (in person) with iterative adjustment of implantable device to test the function of device and to select optimal permanent programmed values, with analysis, review, and report, by a physician or other qualified health care professional, leadless pacemaker system in dual cardiac chambers

 

 

ICD-10 Procedure

 

02HK3NZ

Insertion of intracardiac pacemaker into right ventricle, percutaneous approach

02PA3NZ

Removal of intracardiac pacemaker from heart, percutaneous approach

X2H63V9

Insertion of dual-chamber intracardiac pacemaker into right atrium, percutaneous approach, new technology group 9

X2HK3V9

Insertion of dual-chamber intracardiac pacemaker into right ventricle, percutaneous approach, new technology group 9

 

 

ICD-10 Diagnosis

 

 

All diagnoses

References

Peer Reviewed Publications:

  1. Arkles J, Cooper J. The emerging roles of leadless devices. Curr Treat Options Cardiovasc Med. 2016; 18(2):14.
  2. Breatnach CR, Dunne L, Al-Alawi K, et al. Leadless Micra pacemaker use in the pediatric population: Device implantation and short-term outcomes. Pediatr Cardiol. 2020; 41(4):683-686.
  3. Cantillon DJ, Dukkipati SR, Ip JH, et al. Comparative study of acute and mid-term complications with leadless and transvenous cardiac pacemakers. Heart Rhythm. 2018; 15(7):1023-1030.
  4. Chami El M, Kowal RC, Soejima K, et al. Impact of operator experience and training strategy on procedural outcomes with leadless pacing: Insights from the Micra Transcatheter pacing study. Pacing Clin Electrophysiol. 2017; 40(7):834-842.
  5. Chieng D, Lee F, Ireland K, Paul V. Safety and efficacy outcomes of combined leadless pacemaker and atrioventricular nodal ablation for atrial fibrillation using a single femoral puncture approach. Heart Lung Circ. 2020; 29(5):759-765.
  6. Da Costa A, Axiotis A, Romeyer-Bouchard C, et al. Transcatheter leadless cardiac pacing: The new alternative solution. Int J Cardiol. 2017; 227:122-126.
  7. Darlington D, Brown P, Carvalho V, et al. Efficacy and safety of leadless pacemaker: A systematic review, pooled analysis and meta-analysis. Indian Pacing Electrophysiol J. 2022; 22(2):77-86.
  8. Duray GZ, Ritter P, El-Chami M, et al. Long-term performance of a transcatheter pacing system: 12-month results from the Micra Transcatheter Pacing Study. Heart Rhythm. 2017; 14:702-709.
  9. El-Chami MF, Al-Samadi F, Clementy C, et al. Updated performance of the Micra transcatheter pacemaker in the real-world setting: A comparison to the investigational study and a transvenous historical control. Heart Rhythm. 2018; 15(12):1800-1807.
  10. El-Chami MF, Johansen JB, Zaidi A, et al. Leadless pacemaker implant in patients with preexisting infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019; 30(4):569-574.
  11. Garg A, Koneru JN, Fagan DH, et al. Morbidity and mortality in patients precluded for transvenous pacemaker implantation: Experience with a leadless pacemaker. Heart Rhythm. 2020; 17(12):2056-2063.
  12. Knops RE, Tjong FV, Neuzil P, et al. Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial. J Am Coll Cardiol. 2015; 65(15):1497-1504.
  13. Kypta A, Blessberger H, Kammler J, et al. Leadless cardiac pacemaker implantation after lead extraction in patients with severe device infection. J Cardiovasc Electrophysiol. 2016; 27(9):1067-1071.
  14. Lakkireddy D1, Knops R2, Atwater B3, et al. A worldwide experience of the management of battery failures and chronic device retrieval of the Nanostim leadless pacemaker. Heart Rhythm. 2017; 14(12):1756-1763.
  15. Lenarczyk R, Boveda S, Mansourati J, et al. Peri-procedural management, implantation feasibility, and short-term outcomes in patients undergoing implantation of leadless pacemakers: European snapshot survey. Europace. 2020; 22(5):833-838.
  16. Martínez-Sande JL, García-Seara J, Rodríguez-Manero M, et al. The Micra leadless transcatheter pacemaker. Implantation and mid-term follow-up results in a single center. Rev Esp Cardiol (Engl Ed). 2017; 70(4):275-281.
  17. Miller MA, Neuzil P, Dukkipati SR, Reddy VY. Leadless cardiac pacemakers: Back to the future. J Am Coll Cardiol. 2015; 66(10):1179-1189.
  18. Ngo L, Nour D, Denman RA, et al. Safety and efficacy of leadless pacemakers: A systematic review and meta-analysis. J Am Heart Assoc. 2021; 10(13):e019212.
  19. Okabe T, El-Chami MF, Lloyd MS, et al. Leadless pacemaker implantation and concurrent atrioventricular junction ablation in patients with atrial fibrillation. Pacing Clin Electrophysiol. 2018; 41(5):504-510.
  20. Oliveira SF, Carvalho MM, Adao L, Nunes JP. Clinical outcomes of leadless pacemaker: A systematic review. Minerva Cardioangiol. 2020 Jul 10 [Epub ahead of print].
  21. Piccini JP, Cunnane R, Steffel J, et al. Development and validation of a risk score for predicting pericardial effusion in patients undergoing leadless pacemaker implantation: experience with the Micra transcatheter pacemaker. Europace. 2022; 24(7):1119-1126.
  22. Piccini JP, El-Chami M, Wherry K, et al. Contemporaneous comparison of outcomes among patients implanted with a leadless vs transvenous single-chamber ventricular pacemaker. JAMA Cardiol. 2021; 6(10):1187-1195.
  23. Piccini JP, Stromberg K, Jackson KP, et al. Long-term outcomes in leadless Micra transcatheter pacemakers with elevated thresholds at implantation: Results from the Micra Transcatheter Pacing System Global Clinical Trial. Heart Rhythm. 2017; 14(5):685-691.
  24. Reddy VY, Exner DV, Cantillon DJ, et al. Percutaneous implantation of an entirely intracardiac leadless pacemaker. N Engl J Med. 2015; 373(12):1125-1135.
  25. Reddy VY, Knops RE, Sperzel J, et al. Permanent leadless cardiac pacing: results of the LEADLESS trial. Circ. 2014; 129(14):1466-1471.
  26. Reddy VY1, Miller MA2, Knops RE2, et al. Retrieval of the leadless cardiac pacemaker: a multicenter experience. Circ Arrhythm Electrophysiol. 2016; 9(12).
  27. Reynolds D, Duray GZ, Omar R, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016; 374(6):533-541.
  28. Ritter P, Duray GZ, Steinwender C, et al. Micra Transcatheter Pacing Study Group. Early performance of a miniaturized leadless cardiac pacemaker: The Micra Transcatheter Pacing Study. Eur Heart J. 2015; 36(37):2510-2519.
  29. Roberts PR, Clementy N, Samadi AF, et al. A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry. Heart Rhythm. 2017; 14(9):1375-1379.
  30. Sperzel J, Burri H, Gras D, et al. State of the art of leadless pacing. Europace. 2015; 17(10):1508-1513.
  31. Sperzel J, Defaye P, Delnoy PP, et al. Primary safety results from the LEADLESS observational study. Europace. 2018; 20(9):1491-1497.
  32. Tjong FVY, Knops RE, Neuzil P, et al. Midterm safety and performance of a leadless cardiac pacemaker: 3-year follow-up to the LEADLESS Trial (Nanostim Safety and Performance Trial for a Leadless Cardiac Pacemaker System). Circ. 2018; 137(6):633-635.
  33. Tjong FV, Reddy VY. Permanent leadless cardiac pacemaker therapy: A comprehensive review. Circ. 2017; 135(15):1458-1470.
  34. Vamos M, Erath JW, Benz AP, et al. Incidence of cardiac perforation with conventional and with leadless pacemaker systems: A systematic review and meta-analysis. J Cardiovasc Electrophysiol. 2017; 28(3):336-346.
  35. Yarlagadda B, Turagam MK, Dar T, et al. Safety and feasibility of leadless pacemaker in patients undergoing atrioventricular node ablation for atrial fibrillation. Heart Rhythm. 2018; 15(7):994-1000.
  36. Zucchelli G, Barletta V, Bongiorni MG. Leadless technology: A new paradigm for cardiac pacing? Minerva Cardioangiol. 2018; 66(1):113-123.

Government Agency, Medical Society, and Other Authoritative Publications:

  1. American Heart Association. Statement of the American Heart Association to the Food and Drug Administration Circulatory System Devices Panel February 18, 2016: Leadless Cardiac Pacemaker Devices. 2016. Available at:  https://www.fda.gov/advisory-committees/circulatory-system-devices-panel/2016-meeting-materials-circulatory-system-devices-panel. Accessed on January 18, 2023.
  2. Boveda S, Marijon E, Lenarczyk R, et al. Factors influencing the use of leadless or transvenous pacemakers: Results of the European Heart Rhythm Association prospective survey. Europace. 2020; 22(4):667-673.
  3. Centers for Medicare and Medicaid Services (CMS). National Coverage Determination: Leadless Pacemakers (CAG-00448N). Effective date: January 18, 2017. Available at: https://www.cms.gov/medicare-coverage-database/details/nca-decision-memo.aspx?NCAId=285. Accessed on January 18, 2023.
  4. El-Chami MF, Bockstedt L, Longacre C, et al. Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up. Eur Heart J. 2022; 43(12):1207-1215.
  5. El-Chami MF1, Clementy N2, Garweg C, et al. Leadless pacemaker implantation in hemodialysis patients: experience with the Micra transcatheter pacemaker. JACC Clin Electrophysiol. 2019; 5(2):162-170.
  6. El-Chami MF, Garweg C, Iacopino S, et al. Leadless pacemaker implant, anticoagulation status, and outcomes: results from the micra transcatheter pacing system post-approval registry. Heart Rhythm. 2022; 19(2):228-234.
  7. El-Chami MF, Shinn T, Bansal S, et al. Leadless pacemaker implant with concomitant atrioventricular node ablation: Experience with the Micra transcatheter pacemaker. J Cardiovasc Electrophysiol. 2021; 32(3):832-841.
  8. Han JJ. The Aveir leadless pacing system receives FDA approval. Artif Organs. 2022; 46(7):1219-1220.
  9. Medtronic, Inc. Micra Transcatheter Pacing System Post-Approval Registry. NCT02536118. Last updated February 3, 2022. Available at: https://clinicaltrials.gov/ct2/show/NCT02536118. Accessed on January 18, 2023.
  10. Reddy VY, Exner DV, Doshi R, et al; LEADLESS II Investigators. Primary results on safety and efficacy from the LEADLESS II -- Phase 2 worldwide clinical trial. JACC Clin Electrophysiol. 2022; 8(1):115-117.
  11. U.S. Food and Drug Administration (FDA) Premarket Approval. Micra Transcatheter Pacemaker System.  Summary of Safety and Effectiveness. No. P150033. Rockville, MD: FDA. April 6, 2016. Available at: https://www.accessdata.fda.gov/cdrh_docs/pdf15/P150033B.pdf. Accessed on January 18, 2023.
  12. U.S. Food and Drug Administration (FDA) Summary of Safety and Effectiveness data (SSED). AveirVR Leadless System (Abbot Medical, Sylmar, CA). No. P150035, Rockville, MD: FDA. March 31, 2022. Available at: https://www.accessdata.fda.gov/cdrh_docs/pdf15/P150035B.pdf. Accessed on January 18, 2023.
  13. Zou F, Di Biase L. 2021 European Society of Cardiology (ESC). Guidelines on cardiac pacing and cardiac resynchronization therapy: highlights and major changes. Nov 23, 2021. Available at: https://www.acc.org/Latest-in-Cardiology/Articles/2021/11/23/19/45/ESC-Guidelines-on-Cardiac-Pacing-and-CRT-esc-2021. Accessed on January 18, 2023. 
Websites for Additional Information
  1. Cleveland Clinic Heart failure member information. Available at:  https://pages.clevelandclinic.org/heart-failure-index-2.html?utm_source=bing_ppc&utm_medium=cpc&utm_campaign=Heart+-+Heart+Failure+-+Diagnosis+-+Region+1,+2,+3&utm_term=heart%20failure%20symptoms&gclsrc=aw.ds&msclkid=ae094964d2791b3e02332aacc8fa36dd. Accessed on January 18, 2023.
  2. Medical Device and Diagnostic Industry (MDDI). What is the impact of St. Jude’s voluntary pause of Nanostim implants?  Issued October 28, 2016. Available at: https://www.mddionline.com/what-impact-st-judes-voluntary-pause-nanostim-implants. Accessed on January 18, 2023.
  3. National Heart, Lung and Blood Institute. Conduction disorders/Heart block. Available at: https://www.nhlbi.nih.gov/health/health-topics/topics/hb. Accessed on January 18, 2023.
Index

Aveir single-chamber (VR) leadless pacemaker
Leadless Pacemaker
Micra Transcatheter Pacemaker System, TPS
MR Conditional Pacer
Nanostim Leadless Pacing System
TPS

The use of specific product names is illustrative only. It is not intended to be a recommendation of one product over another, and is not intended to represent a complete listing of all products available.

Document History

Status

Date

Action

  09/27/2023 Updated Coding section with 10/01/2023 ICD-10-PCS changes, added X2H63V9, X2HK3V9.
  06/28/2023 Updated Coding section with 07/01/2023 CPT changes, added 0795T, 0796T, 0797T, 0798T, 0799T, 0800T, 0801T, 0802T, 0803T, 0804T; also added ICD-10-PCS code 02PA3NZ.

Reviewed

02/16/2023

Medical Policy & Technology Assessment Committee (MPTAC) review.Updated Background, Index and References sections.

Reviewed

02/17/2022

MPTAC review. References were updated.

Reviewed

02/11/2021

MPTAC review. The Background and References sections were updated.

Reviewed

02/20/2020

MPTAC review. References were updated. Updated Coding section; added ICD-10-PCS code 02HK3NZ.

 

12/31/2019

Updated Coding section with 01/01/2020 CPT changes; revised descriptor for 33275.

Reviewed

03/21/2019

MPTAC review. References were updated.

 

12/27/2018

Updated Coding section with 01/01/2019 CPT changes; added 33274, 33275; 0387T-0391T deleted 12/31/2018.

Reviewed

03/22/2018

MPTAC review. The document header wording was updated from “Current Effective Date” to “Publish Date.” References were updated.

New

05/04/2017

MPTAC review. Initial document development.


Federal and State law, as well as contract language, including definitions and specific contract provisions/exclusions, take precedence over Medical Policy and must be considered first in determining eligibility for coverage. The member’s contract benefits in effect on the date that services are rendered must be used. Medical Policy, which addresses medical efficacy, should be considered before utilizing medical opinion in adjudication. Medical technology is constantly evolving, and we reserve the right to review and update Medical Policy periodically.

No part of this publication may be reproduced, stored in a retrieval system or transmitted, in any form or by any means, electronic, mechanical, photocopying, or otherwise, without permission from the health plan.

© CPT Only - American Medical Association